Specifiche tecniche | |||||||
---|---|---|---|---|---|---|---|
Scala pH | -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH | ||||||
Risoluzione pH | 0.1 pH / 0.01 pH / 0.001 pH | ||||||
Accuratezza pH | ±0.1 pH / ±0.01 pH / ±0.002 pH ± 1LSD | ||||||
Calibrazione pH | Up to five-point calibration, eight standard buffers available (pH 1.68, 3.00, 4.01, 6.86, 7.01,9.18, 10.01, 12.45), and five custom buffers | ||||||
Compensazione della Temperatura pH | automatic or manual from -20.0 to 120.0 °C | ||||||
Scala mV | ±2000.0 mV | ||||||
Risoluzione mV | 0.1 mV | ||||||
Accuratezza mV | ±0.2 mV ± 1LSD | ||||||
Offset mV Relativi | ±2000.0 mV | ||||||
Scala ISE | e.g. 10-7 to 10M/ 0.005 to 105 ppm/ 5·10–7 to 5·107 conc | ||||||
Risoluzione ISE | 1 conc./ 0.1 conc. / 0.01 conc. / 0.001 conc | ||||||
Accuratezza ISE | ±0.5% (monovalent ions); ±1% (divalent ions) | ||||||
Calibrazione ISE | Up to five-point calibration, seven fixed standard solutions available for each measurement unit, and five custom solutions | ||||||
Scala EC | 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* | ||||||
Risoluzione EC | 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm | ||||||
Accuratezza EC | ±1% of reading (±0.01 μS/cm) | ||||||
Calibrazione EC | automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration | ||||||
Scala TDS | 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) | ||||||
Risoluzione TDS | 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt | ||||||
Accuratezza TDS | ±1% of reading (±0.01 ppm) | ||||||
Compensazione della Temperatura EC/TDS | Disabled / Linear / Non linear (natural water) | ||||||
Coefficiente di Correzione della Temperatura | 0.00 to 10.00 %/°C | ||||||
Temperatura di Riferimento | 5.0 °C to 30.0 °C | ||||||
Costante di Cella | 0.0500 to 200.00 | ||||||
Tipo di Cella | 4 cells | ||||||
Scala Salinità (% NaCl) | 0.0 to 400.0% | ||||||
Risoluzione Salinità (% NaCl) | 0.01% | ||||||
Accuratezza Salinità (% NaCl) | ±1% of reading | ||||||
Scala Salinità (PSU) | 0.00 to 42.00 psu | ||||||
Risoluzione Salinità (PSU) | 0.01 | ||||||
Accuratezza Salinità (PSU) | ±1% of reading | ||||||
Calibrazione Salinità (PSU) | one-point | ||||||
Scala Salinità (ppt) | 0.00 to 80.00 ppt | ||||||
Risoluzione Salinità (ppt) | 0.01 | ||||||
Accuratezza Salinità (ppt) | ±1% of reading | ||||||
Scala Resistività | 1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm | ||||||
Risoluzione Resistività | 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* | ||||||
Accuratezza Resistività | ±2% of reading (±1 Ω•cm) | ||||||
Scala Temperatura | -20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K** | ||||||
Risoluzione Temperatura | 0.1 °C, 0.1 °F, 0.1 K | ||||||
Accuratezza Temperatura | ±0.2 °C; ±0.4 °F; ±0.2 K (without probe) | ||||||
Canali di Ingresso | 2 (pH/mV/ISE; Conductivity/Resistivity/TDS/ Salinity) | ||||||
Elettrodo pH | HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included) | ||||||
Sonda EC/TDS | HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) | ||||||
Sonda Temperatura | HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) | ||||||
Tipo di Registrazione | Automatic, Manual, AutoHold | ||||||
Memoria di Registrazione | Up to 100 lots, 50,000 records max/lot; maximum 100,000 data points / channel | ||||||
Intervallo di Registrazione | 14 selectable between 1 second and 180 minutes | ||||||
Collegamento | Opto-isolated USB | ||||||
GLP | yes (GLP Channel 1: Electrode offset / slope, calibration points, calibration time stamp, GLP Channel 2: Probe cell constant / offset, reference temperature, compensation coefficient, calibration points, calibration time stamp) | ||||||
Avviso Scadenza Calibrazione | yes | ||||||
Conformità USP | USP <645> Application | ||||||
Profili | Up to 10 (5 for each) | ||||||
Display | Color Graphic LCD 240 x 320 pixels, 8 keys capacitive touch keyboard | ||||||
Alimentazione | 12 VDC adapter (included) | ||||||
Condizioni di Utilizzo | 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing | ||||||
Dimensioni | 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) | ||||||
Peso | 1.2 Kg (2.6 lbs) | ||||||
Informazioni Ordine |
HI5522 è fornito con elettrodo pH HI1131, sonda di temperatura HI7662-T, sonda di conducibilità HI76312, portaelettrodo HI76404W, bustina di soluzione tampone HI70004 pH 4.01, bustina di soluzione tampone HI70007 pH 7.01, bustina di soluzione tampone pH 10.01, bustina di soluzione detergente per elettrodi HI700601 (2), elettrodo HI7082 3.5M KCl te soluzione (30 ml), alimentatore da 12 V CC e istruzioni. |